Netzsch LFA447 Nanoflash System

  • Conform to ASTM E1461
  • Measuring thermal diffusivity directly
  • Measuring specific heat with reference sample
  • Thermal conductivity = thermal diffusivity x density x specific heat
  • Xenon flash lamps
  • InSb IR Detector
  • Temperature Range: ambient - 300 C
  • Thermal diffusivity range: 0.001 - 10 cm^2/s
  • Preferred specimen size: round (25.4 or 12.7 mm in diameter) or square (8 mm side length) specimens, up to 3 mm in thickness
  • Accuracy: 5%
  • Repeatability: 3%

Manufacture Wesite