Instrumentation
Mass Spectrometers
Listed below are the current major instrumentation in MUAL, along with key specifications and primary application.
Liquid Chromatography- Mass Spectrometry
Thermo Orbitrap Fusion™ Tribrid™ Mass Spectrometer

- Combines quadrupole, ion trap and Orbitrap mass analysis in a revolutionary Tribrid architecture
- Ultrahigh resolution up to 500,000 FWHM
- Sub-1-ppm mass accuracy
- Multiple dissociation techniques—CID, HCD, ETD
- ID-X capability for small molecule analysis
- Coupled to ESI and nano-ESI interface
- UltiMate 3000 RS UHPLC & UltiMate 3000 RSLCnano
- Primary application: Global metabolomics, Proteomics, Characterization of unknowns
Shimadzu LCMS-8040

- Triplequadrupole mass spectrometer
- Fast scan speeds with ultrafast polarity switching
- 15,000 u/second with a 0.1 Dalton step width without the loss of mass accuracy.
- Redesigned pseudo-quadrupole collision cell, allows for 1msec pause times to eliminate crosstalk
- Coupled to ESI interface
Waters Xevo TQ-XS

- Triple quadrupole mass spectrometer
- Equipped with StepWave XS™ ion guide to enhance sensitivity, T-Wave and ScanWave Collision Cell for rapid simultaneous MS/MS and MS scanning, and Xtended Dynamic Range (XDR) photomultiplier detector allows for larger ranges of linearity.
- Ultrafast scan speeds (20,000 Da/second) with rapid polarity switching.
- Coupled to ESI interface
- MRM tuning with ESI infusion for unmatched sensitivity
- Primary application: Trace analysis, targeted metabolomics, Quantitative proteomics, Quantitative analysis
Gas Chromatography- Mass Spectrometry
Agilent 7250 GC/QTOF

- Quadrupole Time-of-flight mass spectrometer
- Delivers full-spectrum, high-resolution, accurate-mass data
- Low-energy EI for softer ionization and molecular ion enhancement
- Elucidates chemical structures with MS/MS capabilities
- TOF mass accuracy- < 2ppm RMS
- TOF Resolution > 25,000 at m/z 271.896
- Data acquisition rates of up to 50 spectra per second
- Electron Ionization, settable 5-200 eV
- Primary application: Metabolomics, Characterization & quantitation of GC-amenable compounds
Agilent 5975C Series GC/MSD System

- Single quadrupole mass spectrometer
- Standard Electron Ionization
- Triple-Axis Detector for superior sensitivity and spectral integrity
- Coupled to a Pyroprobe 5150 Pyrolyzer
- Can sample solid samples
- Ambient to 1400°C and programmable in 1°C increments
- Cryofocusing to produces sharper peaks
- Primary application: Metabolomics, Characterization & quantitation of GC-amenable compounds
Agilent 240 GC/MS Ion Trap System

- Ion Trap mass spectrometer
- Standard Electron Ionization
- Features MS/MS and MSn capabilities
- Coupled to a Pyroprobe 5150 Pyrolyzer
- Selective ion trapping for higher sensitivity
- Full spectrum scan without any ion loss
- Off-campus
- Primary application: Characterization & quantitation of GC-amenable small molecules
Chromatographs
Liquid Chromatographs
Thermo UltiMate 3000 UPLC; Thermo UltiMate™ 3000 RSLCnano; Waters ACQUITY UPLC I-Class PLUS; Shimadzu UFLC




- Coupled to mass specs and to diode-array and florescence detectors.
Shimadzu 2010 plus, Agilent 7890B, 7890A



- Coupled to flame ionization detectors
Other Instruments
Thermo Scientific X-Series 2 ICP-MS; Astoria Pacific Nutrient Analyzer


- Elemental and nutrient analysis